DETERMINATION OF PALLADIUM
DETERMINATION OF PALLADIUM
AND PLATINUM TRACE CONTENT IN
AND PLATINUM TRACE CONTENT IN
WASTES AND TAILINGS OF
WASTES AND TAILINGS OF
CHEMICAL INDUSTRY
CHEMICAL INDUSTRY
Institute of Nuclear Physics,
Institute of Nuclear Physics, Uzbekistan Academy of Science,Uzbekistan Academy of Science, Tashkent
Tashkent
Laboratory Activation analysis of pure materials Laboratory Activation analysis of pure materials
I
Objective
Objective
:
:
Development of high-sensitivity
Development of high-sensitivity
techniques for radiochemical
techniques for radiochemical
neutron activation determination
neutron activation determination
of palladium and platinum in a
of palladium and platinum in a
wastes and
wastes and
slimes
slimes
of the chemical
of the chemical
industry
Conditions and characteristics of
Conditions and characteristics of
techniques
techniques
Pd
msam = 150 mg, tirr = 10 h, f = 1014 neut.sm2/s.
Analytical radionuclide
109Pd (T
1/2 = 13.5 h)
( 9 gamma-lines 40-780 keV)
Pt
msam = 150 mg, tirr = 10 h, f = 1014 neut.sm2/s.
Analytical radionuclide
197Pt (T
1/2 = 20 h)
Distribution coefficient of Pd and Pt in the
Distribution coefficient of Pd and Pt in the
Dowex
Dowex--1 1 -- H HClCl andand Н Н33РОРО4 4 ion exchange systemion exchange system
Ions Concentration of HCl Concentration of H3PO4
1,2 М 3 М 6 М 9 М 12 М 1 М 3 М 5 М 7 М 12 М Pd(II) 1300 500 120 50 32 800 690 730 540 350 Pd(IV) 1600 350 80 55 33
Pt(II) 104 5000 2800 710 480
Pt(IV) 104 3600 1900 870 430 3000 3000 3000 1400 460
From HCl Dowex-1 except Pd and Pt are absorbes about From HCl Dowex-1 except Pd and Pt are absorbes about
15, from H
Distribution coefficient of Pd and Pt in the
Distribution coefficient of Pd and Pt in the
Dowex
Dowex--1 1 -- HNO HNO33 ion exchangeion exchangesystemsystem Element Concentration HNO3
0,1 М 1 М 5 М 10 М 14 М
Pd 35 100 60 10 6
Pt 500 150 80 10
--- Except Pd and PtExcept Pd and Pt are absorbed Au, Re and Hgare absorbed Au, Re and Hg. .
Distribution coefficient of more than 25 elements less Distribution coefficient of more than 25 elements less
1, and of more than 10 elements less than 5
Eluation curve of the impurity elements in the
Eluation curve of the impurity elements in the
system of Dowex-1 - 1 M HNO
system of Dowex-1 - 1 M HNO33
5 10 15 20 25 30 35 40 0 0.2 0.4 0.6
/
Ai A0 V, mlBa, K, Na, Cs, Cd, Ga, Ir, Mn; Ag, Cr, Ni(Co), Cu, Zn, Fe, Sb; As, In, Se, Te, Sc, U, Th, La-Lu;
Distribution profile of
Distribution profile of PtPt (1) and (1) and PdPd (2) on the (2) on the
chromatographic column. h= 5 sm, d = 0,5 sm chromatographic column. h= 5 sm, d = 0,5 sm 0.5 1.0 1.5 2.0 2.5 3.0 0 0.05 0.10 0.15 0.20 2 1 L, см
/
Ai A0Procedure of analysis
Procedure of analysis
The 150 mg of sample together with standards of Pd The 150 mg of sample together with standards of Pd
and Pt was irradiated in the channel of WWR-SM
and Pt was irradiated in the channel of WWR-SM
reactor for 10 h. In a day after irradiation the sample
reactor for 10 h. In a day after irradiation the sample
was dissolved in HCl+HNO
was dissolved in HCl+HNO33 (3:1). The obtained (3:1). The obtained
solution was evaporated to damp salts, the residue was
solution was evaporated to damp salts, the residue was
dissolved in 1M HNO
dissolved in 1M HNO33 and transferred into and transferred into
chromatographic column filled with Dowex-1. Then
chromatographic column filled with Dowex-1. Then
matrix elements were eluated from column with 25-30
matrix elements were eluated from column with 25-30
ml of 1M HNO
ml of 1M HNO33. The Dowex-1 was transferred into the . The Dowex-1 was transferred into the
bottle for counting and
bottle for counting and -activity of Pd and Pt was -activity of Pd and Pt was
The measuring equipment
The measuring equipment
Gamma-spectrometer consisting from HPGe Gamma-spectrometer consisting from HPGe
the detector (efficiency 20 %, resolution 1,7 keV
the detector (efficiency 20 %, resolution 1,7 keV
for 1332,5 keV
for 1332,5 keV 6060Co) and multichannel analyzer Co) and multichannel analyzer
InSpector-1000 (Canberra, USA).
InSpector-1000 (Canberra, USA).
Spectra processed with use of software package Spectra processed with use of software package
Genie-2000.
Conclusion
Conclusion
On the basis of the carried out investigations, On the basis of the carried out investigations,
RNAA techniques for determination of trace
RNAA techniques for determination of trace
contents of Pd and Pt in a industrial wastes and
contents of Pd and Pt in a industrial wastes and
tailings of chemical industries were developed
tailings of chemical industries were developed. .
These techniques are based on ion exchange
These techniques are based on ion exchange
chromatographic concentration of Pd and Pt.
chromatographic concentration of Pd and Pt.
LD 1.10LD 1.10-5-5 % for Pd and 2.10 % for Pd and 2.10-4-4 % mass for Pt. % mass for Pt. S