Performance Verification

………

Introduction

………

This section discusses the tests used to verify performance following repairs or during routine

………

maintenance.

(3)

All tests can be performed without removing the covers of the monitor.

………

If the monitor fails to perform as specified in any test, repairs must correct the problem before

………

the monitor is returned to the user.

………

Performance Tests

(4)

The battery charge and battery performance test should be performed before monitor repairs

………

whenever the battery is suspected as being a source of the problems.

………

All other tests can be used following repairs or during routine maintenance (if required by your

………

local institution).

(5)

Before performing the battery performance test, ensure that the battery is fully charged. (See

………

Battery Charge below).

………

This section is written using factory-set power-up defaults.

………

If your institution has pre configured custom defaults, those values display.

(6)

Battery Charge

………

To fully charge the battery:

………

1. Connect the monitor to an AC power source using the proper power cord.

………

For the A1, use the PS-120V or PS-240V external power supply and power cord.

(7)

2. For the A1, verify that the EXTERNAL POWER indicator is lit.

………

For the A3, verify that the BATTERY CHARGING/AC SOURCE indicator is lit.

………

3. Charge the battery for at least 8 hours.

………

The battery can require a complete discharge/charge cycle to restore its normal capacity,

………

(8)

………

4. To check for a full charge, perform the procedure described in Battery Performance Test on

………

page 12.

(9)

Kaynakça:

Philips A1 and A3 Patient Monitors- Service Guide Service Guide.

Philips Medical Systems

Referanslar

Benzer Belgeler

It ensures safe processes by means of the locking system which does not allow the

This key is used to set the speed and time during programming and to check the RCF value during the operation..

Connect the centrifuge to a correctly grounded power

• At the end of the adjusted time the motor stats to slow down and ‘br’ is seen on the speed

In case of a power cut or any defects, the instrument would be opened manually to be able to get the

Battery If the monitor has not been used for a long period of time, the battery will need.. Note Storing the monitor for a long period without charging the battery can

during repair the complete set of NiBP performance tests described in this service

diagram, power supply, isolated front end, NiBP control, the SpO2 processing module,