A 60-year-old man was admitted with acute posterior myocardial infarc-tion. Coronary a n g i o g r a p h y showed total occlu-sion of the middle segment of the left circumflex artery (Cx) (Fig. A) and primary percutane-ous coronary inter-vention was performed. A JL4 guiding cath-eter was engaged to the left main coronary artery using a 7-Fr sheath in the right femoral artery. The lesion was crossed with a 0.014-inch guide wire and predilated with a 2.0x20-mm balloon, resulting in TIMI 2-3 flow (Fig. B). A 3.0x24-mm paclitaxel-eluting stent was advanced to the lesion, but the proximal part of the Cx could not be crossed because of an acute angle take-off of the Cx and suboptimal guide position. During catheter manipulations, the guiding catheter sprang uncontrollably into the aortic root, pulling the stent-balloon assembly together with the guide wire outside the coronary artery. For better crossing of the lesion, we planned to change the 0.014-inch guide wire with an extra-support guide wire. Upon retrieval of the system, we discovered that the stent was stripped off the delivery balloon. The dislodged stent was found in the abdominal aorta at the suprarenal level. The JL4 guiding catheter was placed in the descending aorta near the stent (Fig. C). A 5-mm coronary snare was advanced through a snare catheter. The middle portion of the stent was grasped and the catheter was advanced to close the snare loop (Fig. D). However, the stent could not be taken into the catheter (Fig.

Ruken Bengi Bakkal Elnur Alizade Zeki Şimşek Nihal Özdemir Department of Cardiology, Kartal Koşuyolu Heart and Research Hospital, İstanbul

Transcatheter retrieval of an embolized stent together with a failed snare

from the descending aorta

Embolize stentin başarısız kıskaç sistemiyle birlikte inen aorttan

transkateter yolla çıkarılması

Türk Kardiyol Dern Arş - Arch Turk Soc Cardiol 2011;39(6):527 doi: 10.5543/tkda.2011.01584

E). Moreover, the snare was trapped between the struts of the stent and could not be reopened. The entire system was then pulled below the level of renal arteries. An 8-Fr sheath was placed in the left femoral artery and another snare loop was used to grasp the edge of the stent. Subsequently, the stent grasped by the snare and the snare complex were pulled out together (Fig. F). Thus, the embolized stent and the snare complex were successfully removed by using a second snare and a larger catheter from the contralateral femoral artery. A C E B D F

Referanslar

Benzer Belgeler

While the initial X-ray sho- wed the catheter fragment lying between the right antecubital re- gion and the axilla (Fig. 1), cardiac catheterization at the second day revealed that

In this article, we report a successful endovascular treatment of sequential saccular aneurysms extending to the hilar area in renal artery of solitary kidney

Coronary artery angiography showing of the right anterior oblique projection showing the left anterior descending artery (LAD), intercoronary connection (con), and coronary

[8] In addition, surgical removal of the stent or a thrombectomy with aspiration can be performed for acute in-stent thrombosis, but this is only suitable for

The fragment was successfully removed transjugularly through a sheath inserted into the right internal jugular vein advanced through vena cava superior, the right atrium and

After a successful coronary artery bypass graft surgery in which the left internal mammary artery was anastomosed to the left anterior descending artery and a saphenous vein graft

We used an 18-gauge longer needle for the same femoral vein just to the level of the distal part of the fragmented sheath, under fluoroscopy guidance.. However, the twisted end

Transcatheter closure of large pulmonary arteriovenous fistula including pulmonary artery to left atrial fis- tula with Amplatzer septal occluder. Catheter Cardiovasc