238

© 2018 Indian Journal of Medical Research, published by Wolters Kluwer - Medknow for Director-General, Indian Council of Medical Research

Indian J Med Res 148, August 2018, pp 238-239

DOI: 10.4103/ijmr.IJMR_1951_16 Quick Response Code:

Clinical Image

An unusual case of guide wire fracture during percutaneous transluminal

coronary angioplasty

Quick Response Code:

A 64 yr old female patient† was referred to the department of Cardiology, Faculty of Medicine, Ufuk University, Ankara, Turkey, with chest pain, in September 2016. A drug-eluting stent (DES) was implanted in the left anterior descending artery (LAD) in another centre a month ago. Coronary angiogram of the left system showed a fractured remnant guide wire extending from the first diagonal ostium to the thoracic aorta (Figs. 1, 2 and Video). It was decided to catch the distal tip of the remnant wire with loop snare catheter and to retrieve the remnant wire by deep-guide catheter wedge technique (Video). The remnant wire fractured by several attempts and turned to a spiral shape in the proximal part of LAD (Fig. 3). Broken wire parts were removed from the circulatory system. Since this technique failed, by passing through the spiral shape fractured remnant wire, a 3.0 mm×18 mm DES was implanted in the LAD urgently. The mass of the

Patient’s consent obtained to publish clinical information and images.

Fig. 1. An angiographic image of the fractured wire shows an

entrapped guide wire in the left anterior descending. The arrow shows proximal tip of the remnant guide wire.

Fig. 2.The angiography shows the fractured wire in the left

anterior descending artery, and the arrows show the extension of the same to the ascending aorta.

Fig. 3. Spiral shape of the remnant fractured guide wire in the

left anterior descending artery.

¹ Video available at ijmr.org.in.

(2)

UZUNGET & KERVANCİOĞLU: GUIDE WIRE FRACTURE DURING PTCA 239

remnant wire was crushed onto the vessel wall (Video). The retained guide wire tip was isolated from the circulation, and post-procedural TIMI (thrombolysis in myocardial infarction) 3 flow was achieved in the LAD (Video). One month later, control coronary angiography was performed and observed normal intracoranary blood flow (Video).

Sezen Baglan Uzunget* & Celal Kervancioğlu

Department of Cardiology, Faculty of Medicine, Ufuk University, Ankara, Turkey

*For correspondence:

sezenbaglan@hotmail.com

Received December 5, 2016

Şekil

Fig. 3. Spiral shape of the remnant fractured guide wire in the  left anterior descending artery.

Referanslar

Benzer Belgeler

Development of guide cath- eter extensions is in part responsible for the recent explosion in chronic total occlusion (CTO) and high-risk complex PCI (6), allowing the operators

Single coronary artery is a relatively rare congenital anomaly of the coronary tree and is commonly associated with other congenital cardiac anomalies such as bicuspid aortic

In this article, we report an 82-year-old female case of low coronary ostium height at high-risk for surgery due to severe aortic stenosis, in whom valve-in-valve

[7-9] The mechanisms of coronary obstruc- tion are the following: [6] native cuspis displacement and an obstructed coronary ostium, a short distance (<10

In our case, we were unable to advance the transcatheter aortic into a suitable position from the ascending aorta to the aortic root because the guidewire was positioned too

Computed tomography angiography demonstrated an abnormal origin and the abnormal course of the right coronary artery between ascending aorta and the main pulmonary

Successful percutaneous coronary intervention with stent implantation in anomalous right coronary arteries arising from the left sinus of valsalva: a report of two cases. Ayalp

During routine fluoroscopy, a short, 0.038-inch guide wire was identified, with its distal end located in the right common iliac artery.. It was erroneously left there