• Sonuç bulunamadı

運用腦室外引流系統與大氣壓力相通方式測得顱內壓之準確性評估

N/A
N/A
Protected

Academic year: 2021

Share "運用腦室外引流系統與大氣壓力相通方式測得顱內壓之準確性評估"

Copied!
2
0
0

Yükleniyor.... (view fulltext now)

Tam metin

(1)

運用腦室外引流系統與大氣壓力相通方式測得顱內壓之準確性評估 The Accuracy and Reproducibility of Using External Ventricular Drainage System to Measure Intracranial Pressure

中文摘要

本研究目的為測試運用腦室外引流系統裝置與大氣壓力相通方式下測量顱內壓 的準確性與再測信度,並探討不同的腦脊髓液比重是否影響此顱內壓測量方法

的準確性。研究對象為62 位裝置有腦室外引流系統之病患。測量方法係利用腦室

外引流系統裝置與監測器二種測量工具,以隨機分配方式先後測量個案平躺、15

度及30 度三種床頭高度之顱內壓,每一種高度做 6 次。二種測量工具所測得之

顱內壓平均值以皮爾森積差相關、配對t 檢定及 Bland-Altman 分析法,比較二種 測量工具之一致性與準確性,結果以Bland-Altman 分析圖呈現。腦室外引流系 統裝置測量顱內壓之再測信度則以皮爾森積差相關及變異係數做分析。

結果顯示二種測量方式所得之平均測量值在床頭高度0 度時,有顯著相關(r

=.93, p < .001),平均差異值(bias)及標準差為 2.0 ±1.6 mmHg,平均差異值 具統計顯著差異(p < .001),一致性界限(bias ± 2SD)從 - 1.2 到 5.2 mmHg。

在床頭抬高15 度時亦有顯著相關(r = .94, p < .001),平均差異值及標準差為 2.1 ± 1.5 mmHg,平均差異值有統計顯著差異(p < .001),一致性界限從 - 0.9 5.2 mmHg。床頭高度 30 度時也是有顯著相關(r = .95, p < .001),平均差異 及標準差為2.0 ± 1.6 mmHg,平均差異值有統計顯著差異(p < .001),一致性 界限從 - 1.2 到 4.9 mmHg。而腦脊髓液比重對腦室外引流系統與監測器測量之平 均值差異的影響,結果顯示無統計顯著相關(P > .05)。

再測信度方面,6 次測量值之間均呈顯著相關(p < .001),具有良好再測信度,

然而由變異係數比較結果發現,在平躺時CV =3.69%,15 度時 CV = 6.62%,30

度時CV = 7.88%,有床頭高度越高,測量變異越大,穩定度越差之情形。

綜合以上結果發現,腦室外引流系統與監測器測量值之間缺乏一致性,而腦脊 髓液比重並不影響此一致性不佳的結果,因此腦室外引流系統裝置與監測器二 種顱內壓測量工具不可互相取代,在使用腦室外引流系統做顱內壓測量時,建 議以再測信度佳、測量變異最少的平躺姿勢做測量。

關鍵字:腦室外引流系統、顱內壓測量、準確性、再測信度

英文摘要

The purpose of this study was to examine the accuracy and reproducibility of intracranial pressure (ICP) measured by the external ventricular drainage (EVD) system. The effect of specific gravity (SpGr) of cerebral spinal fluid (CSF) on the accuracy of the EVD-measured-ICP was also determined. Sixty-two subjects with EVD system were included. The ventriculostomy catheter connected to both the ICP monitor and EVD system was used to measure ICP. ICP was repeatedly measured 6

(2)

times in three positions: head at 0 degree, head up 15 degree and head up 30 degree.

The accuracy of the EVD-measured-ICP was determined using the Pearson’s

correlation, paired t-test, and Bland-Altman method. Analysis of the reproducibility of ICP measurements by EVD system was done with the Pearson’s correlation and coefficient of variation (CV).

The mean of ICP between two measurements was significantly correlated (r = .93, p <

.001) at head 0 degree position, and the bias (mean difference) showed significant differences (p < .001), the standard deviation was 2.0 ± 1.6 mmHg, and the limit of agreement (bias ± 2SD) were between - 1.2 and 5.2 mmHg. Position at head up 15 degree displayed significant correlation (r = .94, p < .001), the bias showed significant differences

(p < .001), and the standard deviation was 2.1 ± 1.5 mmHg, with the limits of

agreement being - 0.9 and 5.2 mmHg. At head up 30 degree, the correlation between measurements was significant (r = .95, p < .001), the bias had significant difference (p

< .001), the standard deviation was 2.0 ± 1.6 mmHg, and the limits of agreement were - 1.2 and 4.9 mmHg. There was no significant correlation between SpGr of CSF and difference of mean ICP by EVD system and monitor measurements (p > .05).

The EVD-measured-ICP had good reproducibility (p < .001). But the CV at head 0 degree was 3.69%, head up 15 degree was 6.62%, and head up 30 degree was 7.88%.

The CV became worse as the head elevated.

In conclusion, the agreement of ICP measurements between the EVD system and monitor was poor, and the SpGr of CSF did not affect this result. Therefore, the EVD system cannot replace the monitor for ICP measurement. The results suggested good measurement at supine position when ICP was measured by the EVD system.

Key words: External ventricular drainage system, Intracranial pressure measurement, Accuracy, Reproducibility

Referanslar

Benzer Belgeler

From the technical perspective, staffs from hospitals, which are belong to the Infectious Disease Control Network, and non- government hospitals tend to have more knowledge on

The mean of ICP between two measurements was significantly correlated (r = .93, p &lt; .001) at head 0 degree position, and the bias (mean difference) showed significant differences

Therefore, sport center could stress on improvement of air quality in these two fields, through efficient population control or higher ventilation, to increase the comfort of

The determinants of indoor air quality in a sports center 中文摘要

are higher significantly before noon than afternoon..When users are more than 80 in multi- function sport court, users of more than 15 in shooting court and users of more than 40

In summary, down-regulation of aortic endothelial gap junctions in hypertensive rats induced by L-NAME is pa rtially recovered by a short-term treatment of atenolol but fully

The study is using PZB model to design Clinical Laboratory High Risk Reminder (HRR) System service quality questionnaires, to measurement clinical medical staffs their expection

EBCPG provides a comprehensive way to assist clinicians in making decision according to the visualized clinical practice guidelines while