Original article

Development and Validation of a New RP-HPLC Method for the Analysis of Triamcinolone in Human Plasma

Selvadurai MURALIDHARAN1*, Vijayan VENUGOPAL2, Jayaraja KUMAR2, Subramani PARASURAMAN3

1 AIMST University, Faculty of Pharmacy, Pharmaceutical Chemistry Unit, MALAYSIA,

2AIMST University, Faculty of Pharmacy, PharmaceuticalTechnology Unit, MALAYSIA,

3AIMST University, Faculty of Pharmacy, Pharmacology Unit, MALAYSIA

A new accurate and sensitive RP-HPLC method has been developed and validated for the analysis of triamcinolone in human plasma for its bioequivalence studies. In the method, 0.5 % triethylamine (TEA) (pH 3.48) and acetonitrile in the ratio of 50:50 v/v was used as mobile phase at 1 mL/min flow rate and the Thermo C18 column was used as stationary phase. Protein precipitation technique was used for plasma extraction and the extraction recovery was achieved as 79.5 % using 5 % of trichloroacetic acid. The drug was eluted at 4.9 min. and no plasma endogenous materials were observed in the retention time. The limit of quantitation (LOQ) of the method was calculated as 100 ng/mL and the linearity was in the range of 0.5-15.0 µg/mL. R2 of the calibration curve was found as 0.999. The proposed method shows a good accuracy, precision and linearity and, it can be used for the bioequivalence studies of triamcinolone.

Key words: Triamcinolone, Human plasma, HPLC, Optimization, Validation

Triamsinolon’un İnsan Plazmasında Analizi İçin Yeni Bir TF-YPSK Yönteminin Geliştirilmesi ve Validasyonu

Biyoeşdeğerlik çalşmaları için triamsinolon’un insan plazmasında analizi için yeni ve doğru bir TF- YPSK yöntemi geliştirilmiş ve valide edilmiştir. Yöntemde; mobil faz olarak 1 mL/dk akış hızında % 0.5 trietilamin (TEA) (pH 3.48) ve asetonitril’in 50:50 v/v oranında karışımı ve sabit faz olarak da Thermo C18 kolonu kullanılmıştır. Plazma ekstraksiyonunda protein çöktürme yöntemi kullanılmış ve % 5 lik trikloroasetik asit kullanılarak % 79.5 ekstraksiyon verimi elde edilmiştir. Yöntemdeki tayin limiti (LOQ) 100 ng/mL olarak hesaplanmıştır ve doğrusallık 0.5-15.0 µg/mL aralığındadır. Kalibrasyon eğrisinin r2 değeri 0.999 olarak bulunmuştur. Önerilen yöntem iyi bir doğruluk, hassasiyet ve doğrusallık göstermektedir ve Triamsinolon‘un biyoeşdeğerlik çalşmalarında kullanılabilir.

Anahtar kelimeler: Triamsinolon, İnsan plazması, YPSK, Optimizasyon, Validasyon

*Correspondence: E-mail: murali23pharm@hotmail.com; Tel: +60164953594

INTRODUCTION

Triamcinolone acetonide [9-Fluoro- 11ß,16a,17, 21-tetrahydroxypregna-1,4-diene- 3,20-dione cyclic 16,17-acetal with acetone (C24H31FO6)] is a long-acting corticosteroid with reported efficacy when given by intravitreal or subtenon injection as a treatment for diabetic macular edema, uveitis,

retinal vein occlusion, and age-related macular degeneration (1-3). No method reported previously estimation of triamcinolone acetonide in plasma by using the protein precipitation technique (PPT) extraction in the literature. The present work was aimed at developing a sensitive High Performance Liquid Chromatographic (HPLC) for determination of triamcinolone acetonide in human plasma. The advantages of present method include single step

(2)

10

optimized extraction procedure using PPT and short run time. The optimization of extraction procedure was carried out by comparing solid phase extraction, and liquid-liquid extraction for recovery and interference. The PPT was selected because it had huge advantages such as less processing time; very less organic solvent consumption, fewer steps and good plasma sample clean up. Only few methods have been described for the determination of TRE from plasma, urine, hair, bovine tissue and synovial and intraocular fluid, such as HPLC (4-10) and gas chromatography (11, 12). UV spectrophotometric methods for its pharmaceuticals have been described in British Pharmacopoeia (13) and United States Pharmacopoeia (14). Best of our knowledge, there has been no reported analysis method for the determination of triamcinolone in human plasma using PPT method. Therefore, the goal of this present study was to develop an HPLC method for the determination of triamcinolone in plasma which can be used for pharmacokinetic applications. No method reported previously estimation of triamcinolone in human blank plasma by using the protein precipitation (PPT) extraction in the literature. The present work was aimed at developing a sensitive High Performance Liquid Chromatographic (RP- HPLC) for determination of triamcinolone in human blank plasma. The advantages of present method include single step optimized extraction procedure using PPT and short run time. The optimization of extraction procedure was carried out by comparing Solid phase extraction, and liquid-liquid extraction for recovery and interference. PPT was selected because it had obvious advantages such as shorter processing time; lesser organic solvent consumption, fewer steps, and good plasma sample clean up.

MATERIALS AND METHOD

The HPLC grade of acetonitrile, Tri-Ethyl Amine (TEA) was obtained from Merck Darmstadt, Germany. The HPLC water obtained from HPLC Millipore unit, USA.

Method

Development of chromatographic condition

The chromatogram was developing initially using separation condition such as mobile phase (0.5% TEA and acetonitrile in the ratio of 10:90 increasing order). Column such as C4, C8, C18 and ODS in that C18 column achieved good separation. The system was used shimadzu UFLC-20 AD model. The optimized chromatographic conditions were optimized using a mobile phase 0.5% TEA, pH- 3.48 and acetonitrile in the ratio of 50:50 at flow rate 1mL/min. the stationary phase was used as thermo C18column.

Standard solution preparation

Triamcinolone standard was prepared by using HPLC grade methanol and water in the ratio 1:1. Initially 100 mg of drug was weighted and transferred into the standard flask; the combined solvent (methanol and water) added and finally made the volume with the same. The stock solution further diluted was used for the analysis. The stock solution was maintained refrigerated at 8ºC.

Extraction method

The sample was extracted using PPT, solid phase extraction and liquid-liquid extraction.

In that solid phase extraction and liquid-liquid extraction achieved less than 60% with more interference in the drug retention time. Protein precipitation technique is achieved 79.5%.

The healthy drug free human volunteer blood withdrawn and separated the plasma and stored at -7ºC. Initially 0.5 mL of drug free plasma added in the 2 mL RIA vial followed by 0.5 mL of protein precipitant and 0.5 mL of 1µg of drug added and vertexes for 2 minutes. The resulting solution was centrifuged at 4000 rpm for 7 min. The supernatant was recentrifuged for 2 min at 4000 rpm. The clear extracted solution is injected to UFLC system and chromatogram is recorded.

Validation

The validation (15) parameter like selectivity of the method was analysis for interference from plasma. The standard curve consisting of eight points was developed.

Quality control samples of LQC (µg/mL), MQC (µg/mL) and HQC (µg/mL) were used to determine the intra and inter-day precision and accuracy of the assay. Peak area ratios of

(3)

11 optimized extraction procedure using PPT and

short run time. The optimization of extraction procedure was carried out by comparing solid phase extraction, and liquid-liquid extraction for recovery and interference. The PPT was selected because it had huge advantages such as less processing time; very less organic solvent consumption, fewer steps and good plasma sample clean up. Only few methods have been described for the determination of TRE from plasma, urine, hair, bovine tissue and synovial and intraocular fluid, such as HPLC (4-10) and gas chromatography (11, 12). UV spectrophotometric methods for its pharmaceuticals have been described in British Pharmacopoeia (13) and United States Pharmacopoeia (14). Best of our knowledge, there has been no reported analysis method for the determination of triamcinolone in human plasma using PPT method. Therefore, the goal of this present study was to develop an HPLC method for the determination of triamcinolone in plasma which can be used for pharmacokinetic applications. No method reported previously estimation of triamcinolone in human blank plasma by using the protein precipitation (PPT) extraction in the literature. The present work was aimed at developing a sensitive High Performance Liquid Chromatographic (RP- HPLC) for determination of triamcinolone in human blank plasma. The advantages of present method include single step optimized extraction procedure using PPT and short run time. The optimization of extraction procedure was carried out by comparing Solid phase extraction, and liquid-liquid extraction for recovery and interference. PPT was selected because it had obvious advantages such as shorter processing time; lesser organic solvent consumption, fewer steps, and good plasma sample clean up.

MATERIALS AND METHOD

The HPLC grade of acetonitrile, Tri-Ethyl Amine (TEA) was obtained from Merck Darmstadt, Germany. The HPLC water obtained from HPLC Millipore unit, USA.

Method

Development of chromatographic condition

The chromatogram was developing initially using separation condition such as mobile phase (0.5% TEA and acetonitrile in the ratio of 10:90 increasing order). Column such as C4, C8, C18 and ODS in that C18 column achieved good separation. The system was used shimadzu UFLC-20 AD model. The optimized chromatographic conditions were optimized using a mobile phase 0.5% TEA, pH- 3.48 and acetonitrile in the ratio of 50:50 at flow rate 1mL/min. the stationary phase was used as thermo C18column.

Standard solution preparation

Triamcinolone standard was prepared by using HPLC grade methanol and water in the ratio 1:1. Initially 100 mg of drug was weighted and transferred into the standard flask; the combined solvent (methanol and water) added and finally made the volume with the same. The stock solution further diluted was used for the analysis. The stock solution was maintained refrigerated at 8ºC.

Extraction method

The sample was extracted using PPT, solid phase extraction and liquid-liquid extraction.

In that solid phase extraction and liquid-liquid extraction achieved less than 60% with more interference in the drug retention time. Protein precipitation technique is achieved 79.5%.

The healthy drug free human volunteer blood withdrawn and separated the plasma and stored at -7ºC. Initially 0.5 mL of drug free plasma added in the 2 mL RIA vial followed by 0.5 mL of protein precipitant and 0.5 mL of 1µg of drug added and vertexes for 2 minutes. The resulting solution was centrifuged at 4000 rpm for 7 min. The supernatant was recentrifuged for 2 min at 4000 rpm. The clear extracted solution is injected to UFLC system and chromatogram is recorded.

Validation

The validation (15) parameter like selectivity of the method was analysis for interference from plasma. The standard curve consisting of eight points was developed.

Quality control samples of LQC (µg/mL), MQC (µg/mL) and HQC (µg/mL) were used to determine the intra and inter-day precision and accuracy of the assay. Peak area ratios of

triamcinolone standard were fit to linear equation and drug concentration in test samples with the same day standard curve samples were calculated using this equation.

For all the curves the correlation coefficients (r2) were never lower than 0.9.

Selectivity

Selectivity of the method described was investigated by screening six different batches of human blank plasma. Under the proposed assay condition triamcinolone and internal standard had a retention time of 5.83min, rest of the peaks were due to the plasma components. Triamcinolone and internal standard were very well resolved under the proposed chromatographic conditions. None of the drug free plasma samples studied in this assay yield endogenous interference at these retention times (Figure1).

Accuracy and precision

The mean percent accuracy of the proposed method was found to be 97.03 ± 2.91%. The precision of the inter day and intra-day assay were measured by the percent coefficient of variation over the concentration range of LOQ, low, middle and high quality control sample of triamcinolone during the course of validation. The results are given in Table 1.

Linearity

The linearity of each calibration curve was determined by plotting the peak area ratio of triamcinolone to internal standard versus

nominal concentration of triamcinolone. For linearity study seven different concentration o f triamcinolone were analyzed (0.5, 1, 2, 4,6, 8 and 15 μg/mL). The peak area response was linear over the concentration range studied.

Each experiment at all concentration was repeated three times on three separate days to

obtain the calibration data. The coefficient of correlation r2 was found to be 0.9999.

Recovery

The mean extraction recoveries of triamcinolone determined over the concentration of 0.5, 4.0 and 15.0 ng/mL were 85.94, 94.59 and 85.71%. For the mean extraction recovery was 88.74 ± 5.06%.

Speed of analysis

The analysis is performed within 15 min in which 5 min for extraction of plasma and 10 min for chromatography.

Stability studies

Short-term and long-term stock solution stability study was evaluated, which proved no significant deviation from normal value when stored at 4°C. The stability of triamcinolone in plasma was determined by measuring concentration change in quality control samples over time. Stability was tested by subjecting the quality controls to three freeze-thaw cycles and compared with freshly prepared quality control samples. As shown in Table 2, the mean concentration of triamcinolone in quality control samples did

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 min

0 50 100 150 200 250

mAU

TRIAMCINOLONE/4.992

Figure 1. Typical chromatogram of blank plasma

(4)

12

not change significantly within the time period under the indicated storage conditions.

Long-term stability study results concluded that triamcinolone is stable in plasma matrix at least for 30 days when stored at −20°C.

System suitability

System suitability test was performed daily before the run of analytical batch to check detector response to analyse. Both the formulations were well tolerated by all the volunteers in both the phases of study.

RESULTS AND DISCUSSION Development of bioanalytical method

optimization of chromatographic condition in human plasma

A Plasma study of the optimized method was carried out in healthy human plasma samples was prepared by adding 0.5 mL plasma to 2.0 mL centrifuge tube and added 0.5 mL of precipitating agent (10% v/v perchloric acid) vortexed for 2 min. The resulting solution was centrifuged at 4000 rpm for 7 min. The supernatant layer was separated and estimated by HPLC. The method has been developed using protein precipitation technique, achieved 79.5%. The supernant solution was filtered and injected in HPLC system.

Selectivity and sensitivity

Endogenous interferences study was carried out using 10% of perchloric acid in protein

precipitation method. The respective retention time of drug is 4.9 (Figure 2). No interference

peak was recorded at the drug election time.

Sensitivity of the method showed in nanogram level. Limit of detection and limit of quantification was recorded 30 and 100 ng/mL, respectively.

Accuracy and precision

The accuracy of the proposed method was found to be 90 % as per FDA guidelines. The precision studies of the inter-and intra-day assay were measured percentage coefficient variation over the concentration of the quality control samples. The range has been selected 0.5 – 15 mcg/mL and r2 achieved 0.999. These results indicate the fitness of bioanalytical method. Over the concentration Vs mill absorbance unit (Table 1).

Recovery

In literature survey shown 70% achieved less than 90%. Our developed method has been achieved 90% (Figure 3). These percentage recovery indicates, more sensitivity when compare other methods such as solid phase extraction, liquid-liquid extraction and literature survey.

Speed of analysis

The optimized aqueous chromatographic condition was achieved less than 5 min and bioanalytical method was achieved less than 15% min. in those 5 minutes for extraction of

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 min

0 50 100 150 200 250 300

mAU

TRIAMCINOLONE/4.982

Figure 2. Typical chromatogram of spiked plasm

(5)

13 samples from plasma, 10 mins for elution of

drug.

Stability Studies Stock solution stability

Stock solutions were prepared by using methanol and water in the ratio of 1:1. From

the stock solution at the time of bioanalytical method validation preferred different quality control samples. All the quality control samples stored at suitable temperature.

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 min

-10 0 10 20 30 40 50 60

mAU

TRIAMCINOLONE/4.926

Figure 3. Typical chromatogram of extracted plasma for recovery

Table 1. Intra- and Inter-day accuracy and precision of triamcinolone in human plasma Standard concentration

(μg/mL) Average calculated

concentration (μg/mL) SD % Accuracy Inter-day (n=3)

1.0 0.96 0.38

6.0 5.71 0.52

15.0 14.86 0.47 98.5%

Intra-day (n=3)

1.0 0.92 0.47

6.0 5.73 0.82

SD-Standard deviation

Table 2. Stability study of triamcinolone Standard concentration

(μg/mL) Average concentration

(μg/mL) SD

Bench top (n=5)

6.0 5.94 0.46

15.0 14.25 0.82

Freeze thaw stability (n=5)

6.0 5.12 0.76

15.0 14.21 0.75

Long term stability (n=5)

6.0 5.63 1.05

15.0 14.14 0.93

SD-Standard deviation

(6)

14

Short and long term

Middle quality control and high quality control samples were prepared at time of method validation and injected in the system and compared with the short term and long term duration (Table 2).

System suitability

System suitability test was performed daily before the run of analytical batch to check detector response to the analyte. Both the formulations were well tolerated by all the volunteers in both the phases of study (Table 3).

Freeze thaw stability

At the time of method validation samples were taken from the deep freeze and kept for thaw. The different quality control samples compared with freshly prepared thaw samples.

CONCLUSION

In summary, HPLC method quantification of triamcinolone in drug free human plasma were developed and validated as per FDA guidelines. This method has shown more significant merits over those previous available methods. This method offers significant advantages over those previously reported, in terms of improved sensitivity and selectivity, faster run time (5 min) and lower sample requirements. Hence, this method may be useful for single and multiple ascending dose studies in animal and human subjects.

The developed method has excellent sensitivity, reproducibility and specificity.

The method has been successfully used to provide the bioavailability and bioequivalent

study of triamcinolone in animal and human plasma. The developed assay showed acceptable precision, accuracy, linearity, stability, and specificity so applicable to clinical tool.

REFERENCES

1. Martidis A, Duker JS, Greenberg PB, Intravitreal triamcinolone for refractory diabetic macular edema, Ophthalmology 109, 920-927, 2002.

2. Antcliff RJ, Spalton DJ, Stanford MR, Intravitreal triamcinolone for uveitic cystoid macular edema: an optical coherence tomography study, Ophthalmology 108, 765- 772, 2001.

3. Greenburg PB, Martidis A, Rogers AH, Duker JS, Reichel E, Intravitreal triamcinolone acetonide for macular oedema due to central retinal vein occlusion, Brit J Ophthalmol 86, 247248, 2002.

4. Huopalahti RP, Henion JD, Application of supercritical fluid extraction and high performance liquid chromatography/mass spectrometry for the determination of some anabolic agents directly from bovine tissue samples, J Liq Chromatog RT 19 (1), 69-87, 1996.

5. Derendorf H, Rohdewald P, Hochhaus G, Mollmann H, HPLC determination of glucocorticoid alcohols, their phosphates and hydrocortisone in aqueous solutions and biological fluids, J Pharmaceut Biomed 4 (2), 197-206,1986.

6. Althaus ZR, Rowland JM, Freeman JP, Separation of some natural and synthetic corticosteroids in biological fluids and tissues by High-Performance Liquid Chromatograph, J Chromatog B: Biomed Appl 227 (1), 11-23, 1982.

Table 3. System suitability studies

S. No. Parameters

1 Theoretical plate/meter 4752 3 Asymmetric factor 0.93 4 LOD (ng/mL) 47 5 LOQ (ng/mL) 90

6 Capacity factor (k) 1.8

7 Asymmetry 1.25

LOD-Limit of detection LOQ-Limit of quantification

(7)

15 7. Cirimele V, Kintz P, Dumestre V, Goullé, JP,

Ludes, B, Identification of ten corticosteroids in human hair by liquid chromatography-ion spray mass spectrometry, Forensic Sci Int 107, 381-388, 2000.

8. Mora P, Eperon S, Felt-Baeyens O, Gurny R, Sagodira S, Breton P, Guex-Crosier Y, Trans- scleral diffusion of triamcinolone acetonide, Curr Eye Res 30, 355-361, 2005.

9. Beer PM, Bakri, SJ, Singh RJ, Liu, W, Peters III, GB, Miller, M, Intraocular concentration and pharmacokinetics of triamcinolone acetonide after a single intravitreal injection, Ophthalmology 110 (4), 681-686, 2003.

10. Gaillard Y, Vayssette F, Pépin G, Compared interest between hair analysis and urinalysis in doping controls results for amphetamines, corticosteroids and anabolic steroids in racing cyclists, Forensic Sci Int 107, 361-379, 2000.

11. Amendola L, Garribba F, Botrè F, Determination of endogenous and synthetic glucocorticoids in human urine by gas chromatography-mass spectrometry following microwave-assisted derivatization, Anal Chim Acta 489, 233-243, 2003.

12. Hubbard WC, Liu MC, Bickel C, Argenti D, Heald D, Schleimer RP, Measurement of low picomolar levels of triamcinolone acetonide in human bronchoalveolar lavage fluid by gas chromatography-electron-capture negative-ion mass spectrometry, Anal Biochem 290(1), 18- 28, 2001.

13. Ganderton D, British Pharmacopoeia, Her majesty stationary office, London, 1998, pp.

1321, 1978.

14. Novitch M,The United States Pharmacopoeia XXIV. Pharmacopoeial Convention Inc, USA, 2000, p. 1685.

15. FDA Guidance for Industry. Bioanalytical Method Validation, US Department of Health and Human Services, Food and Drug Administration,Centre for Drug Evaluation and Research (CDER), Center for Veterinary Medicine (CVM) May, 2000.

Received: 26.02.2015 Accepted: 06.08.2015

(8)

16

Referanslar

Benzer Belgeler

Simultaneous estimation of six anti-diabetic drugs: glibenclamide, gliclazide, glipizide, pioglitazone, repaglinide and rosiglitazone: development of a novel HPLC method

The present investigation was carried out in the view of establishing a simple, precise, economic and accurate isocratic reverse phase HPLC method for simultaneous estimation of

Gemcitabine hydrochloride was used as internal stand- ard (IS) for the experiments of proposed method be- cause (i) it gives a well-defined and symmetrical peak (retention

Objectives: The aim of this study was to develop and optimize a simple, cost-effective, and robust high-performance liquid chromatography (HPLC) method by taking an experimental

Analytical method development and validation for simultaneous estimation of teneligliptin hydrobromide hydrate and metformin hydrochloride from its pharmaceutical dosage form by

The stability of the analytes in human plasma was assessed by analysis of six replicates of quality control samples at low and high concentration levels at room temperature over 24

In order to find out the accuracy of the proposed method, recovery studies were performed by spiking the sample of a capsule with an appropriate amount of a

A new, simple and specific method for simultaneous determination QTP and its two metabolites, 7-OH QTP and QTP-SF, in rat plasma was developed using one step protein