題名:Accuracy of the EasyTouch blood glucose self-monitorning system: a study of 516 cases

作者:蔡銘川

Ken-shwo Dai; Ken-shwo Dai; Ping Ho; Chien-Chih Chen; Wen-Chung PengMing-Chuan Tsai

貢獻者:醫學檢驗暨生物技術學系 上傳時間:2009-10-02T08:52:46Z

摘要:BACKGROUND: Self-monitoring blood glucose device is an important tool for diabetes patients to efficiently

control their blood glucose concentrations. We evaluated the accuracy of EasyTouch glucose monitoring system. METHODS: Capillary blood glucose concentrations measured using EasyTouch and the reference values obtained from Yellow Springs Instruments (YSI) 2300 STAT were

performed in the Department of Laboratory Medicine, Wei-Gong Memorial Hospital. Results were evaluated using (1) linear regression analysis, (2) Clarke Error Grid

analysis, (3) percentage of readings within a defined range of deviation from the reference value, (4) bias plots, and (5) coefficients of variation (CVs)

calculated from 60 measurements in series. RESULTS: The window of the 516 EasyTouch readings covered a range from 42 to 555 mg/dl. Linear regression analysis yielded a regression slope 0.9972, intercept 1.899 mg/dl, r2 0.9571, and Syx 14.89 mg/dl. A Clarke Error Grid analysis showed 100% of the EasyTouch readings in clinically acceptable zones A and B. Of the EasyTouch readings, 98.3%, 91.9%, 78.3% and 46.9% were found

within +/-20%, +/-15%, +/-10%, and +/-5%, respectively, of the reference values. Further analysis showed that the percentage of EasyTouch readings within the defined intervals was similar in three glucose ranges (< or =100, 101-200, and > or =201 mg/dl). The CVs for the four lots of strips (lot 1 to lot 4) ranged from 3.5 to 5.5%, 2.1 to 4.8%, 1.8 to 3.6%, and 3.0 to 5.7%,

(2)

accurate and precise glucose readings over a wide range of glucose concentrations.

Referanslar

Benzer Belgeler

O zamana kadar Turklerde vatan diye bir kavram yok, gogebe bir kavim oldugu igin Turklerde vatan degil yurt vardi.. Yurt da gadir

The encouragement of the function ƒ using any feed forward function approaching neural network framework, such as, a standard multi layer preceptons MLP, an radial

In our study, it was observed that the mean deviation values in the coronal deviation, apical deviation and angular deviation parameters were higher in

2014, non-pregnant pa- tients over the age of 18 who were not diagnosed with diabetes and prediabetes who applied Health Sciences University Sisli Hamidiye Etfal Training and

The 2 h results were similar among groups in terms of age, BMI, multiparity, neonatal hypoglycemia, hyperbilirubinemia, intensive care unit admission, birth weight, abnormal result,

Güzel’in (2006), Ankara ili merkez ilçelerinde, annelere yönelik eğitim programlarının geliştirilmesinde kullanılmak üzere annelerin ihtiyaç duydukları konuları

The first chapter analyzes developments on the political scene; the second focuses more on political happenings within the civil society, with emphasis on the issue

Glucose-sensitive hydrogel based systems are potentially appropriate to the development of self-modulated insulin delivery system, and with the manipulation of chemical