• Sonuç bulunamadı

A Biomechanical Comparison of All-Inside Meniscus Repair Techniques

N/A
N/A
Protected

Academic year: 2021

Share "A Biomechanical Comparison of All-Inside Meniscus Repair Techniques"

Copied!
1
0
0

Yükleniyor.... (view fulltext now)

Tam metin

(1)

A Biomechanical Comparison of

All-Inside Meniscus Repair Techniques

李建和

Chang JH;Shen HC;Huang GS;Pan RY;Wu CF;Lee CH;Chen Q

摘要

Abstract

BackgroundThe aim of this study was to assess the biomechanical characteristics of six all-inside meniscal single suture repair techniques using a porcine model.

Materials and MethodsPeripheral longitudinal tears were created in freshly isolated porcine menisci. Tears were repaired using the single vertical technique with six different repair complexes including those involving sutures (#2 FiberWire, #2 Ethibone, flexible anchors (Fast-Fix, RapidLoc), and rigid anchors (Meniscal-Dart, BioStinger). Displacement, ultimate failure strength, stiffness, and site of failure were measured using a Materials Testing System machine. An initial 2 N preload was applied, followed by loading between 5 and 20 N for 300 cycles. Failure strength was determined lastly by increasing tension at a rate of 5 mm/min until failure.

ResultsFailure strength was highest in the #2 FiberWire group (175.6 N). This was

significantly higher than in all other groups (P < 0.05). The second highest failure load was evident in the #2 Ethibone group (113.8 N). This was significantly higher than in all other groups bar the #2 FiberWire group (P < 0.05). Stiffness was also significantly higher in the #2 FiberWire group compared with all other groups (8.5 N/mm, P < 0.05). There were no between-group differences in displacement. When grouped by repair technique, failure load was significantly higher, and displacement was significantly lower, in suture compared with both flexible and rigid anchor repaired menisci (P < 0.01 for all comparisons). Although stiffness was also higher in the suture group, there were no significant between-group differences detected.

ConclusionsSuture techniques exhibited biomechanical superiority over biodegradable flexible and rigid anchor devices for meniscus repair.

Referanslar

Benzer Belgeler

A possible complication of employing the all-inside suture technique at first stage in repair of these tears is that, following the placement of the first implant behind the

However, in the 3 large developed countries of United States, France, and Germany, where intermediate industry momentum was present, intermediate industry momentum portfolios

B. SYSTEM IMPLEMENTATION: This method has the subsequent modules for the implementation of the system. Registration: This module helps the new users to induce registered with

Based on the percentage analysis result, it is revealed that the level of family dynamics and its dimensions of prospective teachers with regard to type of family

Thermal conductivity in the selected direction of the plate of the heat-shielding composite material depending on the angle of orientation of the fibers and

The incident was classified as a Public Health Emergency of International Concern by the World Health Organization (WHO) on 30 January 2020 and declared a pandemic on 11 March

Also, postgraduate students need strong moral support from all concerned, especially their supervisors, friends, families, and employers, as learning at this high level

學術單位 醫療單位