Correction of a migrated Tenckhoff peritoneal

dialysis catheter using a Lunderquist

guidewire: report of two cases

李志明

Jwo SC;Lee CM;Tsai CJ

摘要

Abstract

Two chronic ambulatory peritoneal dialysis (CAPD) patients who experienced from ultrafiltration failure resulting from malposition of a Tenckhoff catheter were treated. Conservative management such as changing body position, saline infusion, and enemas had been tried, but had failed. To avoid surgical intervention, we attempted to correct the malposition using a 120-cm-long Lunderquist guidewire (Nycomed, NY) with a

15-cm-long soft and flexible tip at its distal end. Under fluoroscopy, we successfully repositioned these two migrated Tenckhoff catheters using a Lunderquist guidewire. To the best of our knowledge, this is the first report of using a Lunderquist guidewire to correct a malfunctioning Tenckhoff catheter in CAPD patients. The Lunderquist guide wire has the advantages of being relatively non-invasive and easily used and it provides a reduced morbidity rate. Moreover, using this guidewire allows the Tenckhoff catheter to produce torque and whiplash, buckling, sweeping and rotating maneuvers that can help to correct malposition of the catheter and redirect the catheter to its ideal position. We therefore suggest that patients who receive surgical revision for a malfunctioning Tenckhoff catheter have at least one attempt at correction using this safe and easy procedure before surgery

Referanslar

Benzer Belgeler

During withdrawal of the catheter, the distal marker of IVUS catheter was separated and this segment was moved toward the PLA IVUS - intravascular ultrasound, PLA -

We report a case of an accidental iatrogenic J-wire migration into vena cava during subclavian vein catheter insertion for central venous pressure monitoring in a woman and

While the initial X-ray sho- wed the catheter fragment lying between the right antecubital re- gion and the axilla (Fig. 1), cardiac catheterization at the second day revealed that

Bu çalışmada farklı etiyolojik nedenlere bağlı serebral derin venöz yapılarında trombüs saptanan ve kranial görüntülemede bilateral talamik ve bazal ganglion

Kateter ile ilişkili tromboz nedeniyle gelişen semptomatik VKSS ise literatürde çok nadir olarak bildirilmektedir.. Olgumuzda kateter takılmasından 15 ay sonra, yani geç

Key words: Hydrocephalus, scrotum, migration, ventriculoperitoneal catheter J Nervous Sys Surgery 2008; 1(3):181-184.. Ventriküloperitoneal Şantın Abdominal Kateterinin

Conclusion: Paramedian is a better approach for PD catheter insertion due to lower early and late complications, as well as a higher success rate.. Keywords: Peritoneal

In the paranasal sinus computerized tomography (CT), the concha bullosa was enlarged, filling the right nasal cavity and shifting the septum to the left and a hypointense soft