Inappropriate sensing events revealing electrocautery-induced implantable cardioverter-defibrillator lead failure

N/A
N/A
Protected

Academic year: 2021

Share "Inappropriate sensing events revealing electrocautery-induced implantable cardioverter-defibrillator lead failure"

Copied!
1
0
0

Yükleniyor.... (view fulltext now)

Tam metin

(1)

E-page Original Images

E-27

Inappropriate sensing events

revealing electrocautery-induced

implantable cardioverter–defibrillator

lead failure

Electromagnetic interference (EMI) associated with current implant-able cardioverter–defibrillator (ICD) generators and leads are less prone to long-term failure. Thus, it has been suggested that routine post-ICD replacement interrogation of the device may not be necessary. We pres-ent the case of a patipres-ent who underwpres-ent ICD replacempres-ent during which therapies were inadvertently not turned off, which lead to EMI and shock and subsequently lead failure.

A 52-year-old-man who had ICD (BiotronikLumax340 VR-T, Berlin, Germany) implanted in 2008 presented with an ICD generator end-of-life and was scheduled for generator replacement. Before the procedure device interrogation showed a right ventricular (RV) pacing threshold of 0.6 V at 0.5 ms pulse width, R-wave sensing was 6.7 mV, RV lead pacing impedance was 715 Ω, and shock impedance was 46 Ω. The patient received a Medtronic D384DRG ICD, Minneapolis, USA. During the surgi-cal closure of the device pocket, electrocautery was used for hemosta-sis, which resulted in EMI and inappropriate 35 J shock (Fig. 1). Postoperatively, inappropriate senses were monitored, which were con-cordant with lead failure (Fig. 2). The RV pacing threshold increased to 1.25 V at 0.5 ms, and the R-wave sensing decreased to 3.30 mV. Lead impedance measurements were RV pacing at 619 Ω, RV coil at 48 Ω, and SVC coil at 73 Ω. At the second week, a new pace sense lead was uneventfully implanted from the same site and over sense completely ended after the replacement of the lead.

Despite the advances in ICD technology, electrocautery, especially when used close to the device, can still lead to lead failure, which might necessitate intervention. Thus, we still recommend routine postproce-dural interrogation of the device.

İlknur Can, Alpay Arıbaş, Yüksel Dereli*, Venkatakrishna Tholakanalli1

Departments of Cardiology and *Cardiovascular Surgery, Meram Medical Faculty, Necmettin Erbakan University; Konya-Turkey

1Department of Cardiology, Minnesota University; Minnesota-USA

Address for Correspondence: Dr. Yüksel Dereli, Necmettin Erbakan Üniversitesi

Meram Tıp Fakültesi, 42080, Meram, Konya-Türkiye Phone: +90 332 223 60 00 E-mail: yuxel.dereli@mynet.com

©Copyright 2015 by Turkish Society of Cardiology - Available online at www.anatoljcardiol.com DOI:10.5152/AnatolJCardiol.2015.6338

Figure 1. ECG monitoring showed electrocautery-induced electromagnetic interference and inappropriate 35J shock.

Figure 2. Postoperatively, inappropriate senses were monitored, which were concordant with lead failure.

Referanslar

Benzer Belgeler

Atrial conduction over both slow and fast AV nodes is an underlying cause of this arrhythmia, which explains the double ventricular response to a single atrial signal.. However,

Although SCAD is a known leading nonatherosclerotic cause of sudden cardiac death (SCD) related to myocardial isch- emia presenting with life-threatening ventricular arrythmias in 3%

Hayes SN, Kim ESH, Saw J, Adlam D, Arslanian-Engoren C, Econ- omy KE, et al.; American Heart Association Council on Peripheral Vascular Disease; Council on Clinical Cardiology;

Lead extraction was accomplished by simple traction for 4 atrial, 1 ventricular, and 1 coronary sinus lead (only test stylet inserted), sole use of the locking stylet for 60

In this study, we observed that 14.6% of patients with AF without known arterial disease and 36.6% of patients with diabe- tes were receiving concomitant aspirin

After delivery, the mother continued her ivabradine treatment, and the ECG of the baby, who was fed only breast milk, showed sinus rhythm with a heart rate of 180 bpm (d)..

Partial extraction was 2.5% and failure was Excimer laser assisted implantable cardioverter defibrillator lead extraction:?. An alternative treatment to

Ressamlarımıza Anadolu şehirlerin­ de bir ay müddetle resim yaptırmak suretile Türk sanat ve sanatkârlarını teşvik ve himaye zımnında gösterdiği­ niz büyük